32-Well Real-Time PCR System
32-Well Real-Time PCR System - ToronPCR™ 320T
The ToronPCR™ 320T is a compact 32-well real-time PCR system built for research labs, teaching facilities, and field-ready testing teams that run smaller sample batches. It combines a 0.2 mL thermal block with four fluorescence detection channels, so each well can quantify up to four targets. A maintenance-free LED light source and fiber optic detection keep the optical path simple and stable. Operators set up and monitor runs on an 8-inch touchscreen without a separate computer.
The 320T accepts 8-well strip tubes and single transparent tubes with reaction volumes from 20 to 120 µL. Its block holds ±0.15°C accuracy and ±0.25°C uniformity, giving repeatable amplification from well to well. Fast ramping reaches up to 5°C/sec, which shortens each cycle and the full run. Detection reaches a single target copy across 10 logs of linear dynamic range.
At 37 × 28 × 25 cm and 10 kg, this quantitative thermal cycler fits on crowded benches and moves easily between rooms. Labs use it for gene expression, genotyping, pathogen detection, and agri-food screening. The ToronPCR™ 320T is intended for research use only and is not for diagnostic procedures.
Applications
The ToronPCR™ 320T serves molecular biology workflows where a full 96-well plate is more capacity than the lab needs. Its four-channel optics support both singleplex screening and multiplex assay work.
- DNA and RNA Quantification: measures nucleic acid copy number in extracts for downstream sequencing, cloning, and assay setup
- Gene Expression Analysis: tracks relative expression changes with reference-gene normalization in cell biology and drug response studies
- Pathogen Detection Research: identifies bacterial and viral targets in research samples using probe-based assays
- Genotyping and SNP Analysis: separates alleles with dual-probe assays read in separate fluorescence channels
- Mutation Screening: flags sequence variants in research models and candidate gene panels
- Molecular Cloning: confirms insert presence and estimates plasmid copy number in transformed clones
- Transgenic Research: detects and quantifies transgene integration in engineered plants and animal models
- Agri-Food Safety: screens food and crop samples for contaminating microbes and genetically modified material
- Veterinary Testing: detects animal pathogens in livestock and companion animal research samples
Key Features
The ToronPCR™ 320T packs multi-channel optics, fast thermal cycling, and standalone operation into a small benchtop footprint. Each feature targets quick setup and dependable quantification for low-to-medium sample volumes.
- Four-Channel Fluorescence Detection: runs up to four targets per well, which reduces reagent cost, sample volume, and consumable use
- Maintenance-Free LED Excitation: delivers stable light output with no lamp replacement or warm-up period
- Fiber Optic Detection Path: carries emitted light directly to photodiodes for consistent signal capture across all wells
- Fast Thermal Ramping: heats at up to 5°C/sec to shorten cycle times and speed up results
- Precise Temperature Control: holds ±0.15°C accuracy and ±0.25°C uniformity for repeatable Ct values
- Single-Copy Sensitivity: picks up very low target levels and resolves 1.5-fold differences in singleplex reactions
- 8-Inch Touchscreen with Standalone Operation: lets users create protocols, start runs, and monitor results with no external PC
- Live Amplification Monitoring: displays amplification curves during the run so users can review data as it develops
- Multiple Analysis Modes: supports absolute and relative quantification workflows within the onboard software
- Compact, Lightweight Design: fits small labs and shared benches at 10 kg and a 37 × 28 cm footprint
- Standard 0.2 mL Consumables: accepts common 8-well strip tubes and single tubes, so labs can use existing supplies
Theory and Method
Real-time PCR (qPCR) follows DNA amplification cycle by cycle, rather than measuring product only at the end. Each cycle denatures double-stranded DNA near 95°C, anneals primers at a lower temperature, and extends new strands with a thermostable polymerase. A fluorescent reporter, either an intercalating dye such as SYBR Green or a hydrolysis probe, emits light in proportion to the amount of product. During the exponential phase, the signal roughly doubles with every cycle.
The instrument records the quantification cycle (Cq or Ct), the point where fluorescence crosses a set threshold. Samples with more starting template reach that threshold sooner. Ct falls linearly as the log of the initial copy number rises, so a standard curve turns Ct values into absolute quantities. The ΔΔCt method compares a target with a reference gene for relative quantification. The 320T supports both approaches across 10 logs of linear range.
The 320T excites each well with LEDs between 475 and 620 nm and routes emitted light through optical fibers to photodiodes covering 515 to 670 nm. Four filter channels separate distinct dyes, so one well can carry an internal control plus up to three additional targets. The thermal block heats at an average of 4°C/sec and cools at 3°C/sec, keeping total run time short.
Technical Specifications
Parameter | Specification |
Model | ToronPCR™ 320T |
Description | 32-well real-time PCR instrument with 4 dye channel filters and 8-inch touchscreen |
Sample Capacity | 32 wells, 0.2 mL block |
Compatible Consumables | 8-well strip tubes, single tubes (transparent) |
Reaction Volume | 20 to 120 µL |
Excitation Source | LED |
Excitation Range | 475 to 620 nm |
Detection Method | Photodiodes |
Detection Range | 515 to 670 nm |
Detection Channels | 4 |
Temperature Range | 4 to 100°C |
Average Heating / Cooling Rate | 4°C/sec / 3°C/sec |
Maximum Heating Rate | 5°C/sec |
Temperature Uniformity | ±0.25°C |
Temperature Accuracy | ±0.15°C |
Linear Dynamic Range | 10 logs |
Detection Sensitivity | 1 copy |
Fold-Change Resolution | Detects 1.5-fold differences in target quantity (singleplex) |
Display | 8-inch touchscreen |
Dimensions (W × D × H) | 370 × 280 × 250 mm |
Weight | 10 kg |