Multi-Stage Cascade Impactor

Product Categories

Multi-Stage Cascade Impactor designed for precise analysis of aerodynamic properties of inhaled fine particles, this advanced system includes a smart pump, pressure detector, cascade impactor, pre-separator (DPI), artificial larynx, and essential accessories. It meets the instrument qualification standards outlined in USP, Ph. Eur., and ChP 2020 for aerodynamic particle assessment.

Multi-Stage Cascade Impactor

Multi-Stage Cascade Impactor

  • United States Pharmacopeia (USP) <601> Aerosols, Nasal Sprays, Metered-Dose Inhalers, and Dry Powder Inhalers
  • European Pharmacopoeia (Ph. Eur.) 2.9.18: Preparations for Inhalation - Aerodynamic Assessment of Fine Particles
  • Chinese Pharmacopoeia (ChP) 2020: Inhalation Product Testing
  • ISO 20072: Aerosol Drug Delivery Device - Methods of In Vitro Testing
  • ICH Guidelines for Inhalation Product Evaluation

  1. Contains 7-stage impactors and a micro-orifice collector (MOC). The design allows easy access to collect samples from collection cups and prepare for the next test.
  2. Features 8 removable collection cups for effortless cleaning and replacement.
  3. Suitable for testing MDI, spray, and nebulizer devices. DPI testing can be conducted with a pre-separator that efficiently removes coarse particles of 10-15 microns.
  4. The collection cup serves as the impact surface, placed on a tray directly beneath the nozzles to facilitate sampling.

StageNozzle Diameter (mm)Cutting Diameter (μm)
15L/min3OL/min60L/min100L/min
Stage 114.30±0.0514.1011.768.066.12
Stage 24.88±0.048.616.404.463.42
Stage 32.185±0.025.393.992.822.18
Stage 41.207±0.013.302.301.661.31
Stage 50.608±0.012.081.360.940.72
Stage 60.323±0.011.360.830.550.40
Stage 70.206±0.010.980.540.340.24
Micro-orifice collection (MOC)about 0.0700.700.360.140.07