Face Mask Clogging Tester ToronFM™ CT The sample is exposed to a known concentration of dolomite dust in the air. After exposure, the breathing resistance and filter penetration of the particle-filtering half mask are measured.

Standards

  1. EN 149 section 8.10

ParameterSpecification
ModelToronFM™ CT
DustDRB 4/15 dolomite
Dust DistributorParticle size range: 0.1 um ~ 10 um
Mass Flow Range: 40mg/h --- 400mg/h
Breathing MachineDisplacement: 2.0L/stroke
Frequency: 15 times/min
Exhaled Air Temperature(37 ÷ 2)°C
Exhaled Air Humidity95% R.H minimum
Continuous Flow Through the Dust Chamber60 m3/h, linear velocity: 4 cm/s
Concentration of the Dust(400 ÷ 100) mg/m3
Test ChamberInner Dimension: 650 × 650 mm × 700 mm
Airflow: 60 m%/h, linear velocity: 4 cm/s
Air Temperature: (23 ÷ 2)°C
Air Humidity: (45 ÷ 15)%
Respiratory Resistance Measurement Range0 ~ 2000 Pa, accuracy: 0.1 Pa
Voltage220V, 50Hz, 1KW (110V is also available)
External Dimension3800mm x 1100mm x 1650mm
Weight320kg

Accessories

ItemQuantity
Automatic dolomite dust generator1pc
Automatic powder feeding center system (Continuous feeding device)1set
TSI laser dust particle counter1pc
Sinusoidal wave simulates ventilator1set
Saturated steam generator1set

Request Quote