Trạm làm việc chiết tách axit nucleic

Nucleic Acid Extraction Workstation - ToronNAE™ 4800

ToronNAE™ 4800 High-Capacity Automated Nucleic Acid Extraction Workstation

The ToronNAE™ 4800 nucleic acid extraction workstation automates every step between the primary sample tube and a loaded PCR plate for up to 48 samples per run. Clinical, blood screening, and public health labs use it to remove manual pipetting from extraction and assay setup. A three-channel pipetting arm transfers samples and reagents, a magnetic bead module purifies the nucleic acid, and the same arm then dispenses PCR or qPCR reactions. The whole process runs without operator intervention once the deck is loaded.

Speed suits high-volume testing. The workstation pipettes 48 samples in about 13 minutes and completes extraction in 15 to 30 minutes, depending on the reagent kit. It processes 50 µL to 1000 µL per sample and elutes 40 µL to 200 µL. Operators can choose bottled reagents or prefilled cartridges, and single-sample consumables are supported for small batches. Purification and PCR setup can run in parallel to shorten turnaround time.

Contamination control runs throughout the enclosure. Air displacement pipetting and anti-drip technology keep liquid away from the pipetting channels, while a negative-pressure system with three HEPA filters traps aerosols. Separate UV lamps cover the extraction, pipetting, and waste areas. A built-in barcode reader tracks every sample and reagent through the run. A modular worktable accepts custom layouts and additional devices.

Nucleic Acid Extraction Workstation - ToronNAE™ 4800 Applications

The ToronNAE™ 4800 serves labs that need traceable, hands-off extraction and PCR setup at 48-sample scale. Its sample rack accepts primary tubes, test tubes, and centrifuge tubes for a broad range of specimen types.

  • Clinical Molecular Diagnostics: automated extraction and PCR setup from swabs, blood, and saliva for infectious disease testing
  • sàng lọc máu: nucleic acid testing of serum, plasma, and whole blood for transfusion-transmitted pathogens
  • Public Health and Epidemic Response: high-volume pathogen testing during outbreaks, with barcode traceability for every sample
  • Animal Husbandry and Veterinary Medicine: pathogen and genotyping workflows on animal tissues, blood, and swabs
  • Kiểm tra an toàn thực phẩm: extraction from microorganisms and cell culture fluids for foodborne pathogen and GMO assays
  • Vi sinh môi trường: nucleic acid recovery from soil and environmental microorganisms
  • Phòng thí nghiệm pháp y: consistent DNA extraction from buccal swabs, blood, and tissue for identification work
  • Pathology and Research Labs: DNA extraction from FFPE tissue and purification of PCR or enzyme reaction products
Trạm làm việc chiết tách axit nucleic

Nucleic Acid Extraction Workstation - ToronNAE™ 4800 Key Features

The ToronNAE™ 4800 combines liquid handling, magnetic bead extraction, and PCR setup on one modular deck. Each feature targets faster turnaround, full traceability, and lower contamination risk.

  • Sample-to-PCR Automation: pipetting, lysis, washing, elution, and PCR or qPCR setup run in one workstation without manual transfers
  • Fast 48-Sample Runs: about 13 minutes to pipette 48 samples and 15 to 30 minutes for extraction, depending on the kit
  • Three-Channel Pipetting Arm: two 1000 µL channels and one 200 µL channel cover 1 µL to 1000 µL with CV as low as 0.4%
  • Pressure-Based Liquid Detection: senses liquid level, tip clogs, and residual liquid on every aspiration
  • Liquid-Specific Parameters: built-in liquid libraries and adjustable settings improve accuracy for each reagent type
  • Parallel Purification and PCR Setup: runs extraction and reaction setup side by side to shorten turnaround time
  • Khả năng truy xuất nguồn gốc bằng mã vạch: automatic tube scanning and external USB scanner support track samples and reagents through the run
Trạm làm việc chiết tách axit nucleic

Vortex Magnetic Beads and PCR Reagent Block

Trạm làm việc chiết tách axit nucleic

High-Precision Pipetting

Trạm làm việc chiết tách axit nucleic

Effective Cross-Contamination Prevention

Trạm làm việc chiết tách axit nucleic

Fast & Flexible Experiment

  • Flexible Sample Rack: 48-position rack accepts sample tubes, test tubes, and centrifuge tubes, including primary tubes
  • Bottled or Prefilled Reagents: choose open bottled reagents or pre-packaged cartridges, with single-sample consumables supported
  • Magnetic Bead Vortex: stable vortexing keeps bead stocks uniform for consistent binding across all samples
  • Heated Extraction Area: built-in incubator heats lysis and elution steps up to 120 °C
  • Optional Cooling Block: 0 °C to 105 °C block keeps PCR reagents cold, with customizable reagent block formats
  • Air Displacement Pipetting and Anti-Drip Control: prevents liquid contact with channels and drip contamination across the deck
  • Three-Zone UV and HEPA Protection: separate UV lamps for extraction, pipetting, and waste areas plus three HEPA filters under negative pressure
  • Smart Waste Container: calculates tip fill level and expands storage space for used tips
  • Graphical Software with Consumable Libraries: build complete protocols without complex programming using built-in carrier and liquid libraries
  • Modular Worktable: customizable deck layouts and integration of additional devices as needs grow

Lý thuyết và Phương pháp

Magnetic bead extraction isolates nucleic acids on the surface of silica-coated paramagnetic particles. A lysis buffer with chaotropic salts breaks open cells and viral particles and strips proteins from DNA and RNA. Under these conditions, nucleic acids bind tightly to the bead surface. Wash buffers remove salts, proteins, and PCR inhibitors while the nucleic acids stay attached. A low-salt elution buffer, heated in the extraction area, then releases purified DNA or RNA into solution.

The ToronNAE™ 4800 links this chemistry to a liquid handler. The arm first reads each tube barcode and transfers samples and reagents into the extraction plate. Magnetic rods inside a disposable comb carry the beads through lysis, wash, and elution wells, while a heated incubator holds lysis and elution steps at up to 120 °C. A dedicated vortex keeps the bead stock evenly suspended for uniform dosing. Once elution finishes, the arm dispenses eluate and master mix into PCR plates or strips.

Air displacement pipetting keeps a cushion of air between the piston and the liquid, so samples never touch the pipetting channel. Pressure sensing on each channel detects the liquid surface, residual volume, and tip clogs during aspiration. Liquid-class parameters adjust speed, delay, and blowout for viscous or volatile reagents, which raises accuracy across reagent types. Anti-drip control holds liquid in the tip as the arm crosses the deck. Together these controls keep extraction uniformity at a CV of 5% or less.

Nucleic Acid Extraction Workstation - ToronNAE™ 4800 Technical Specifications

Tham sốĐặc điểm kỹ thuật
Chức năngAutomatic sample pipetting, nucleic acid extraction, and PCR/qPCR setup
Phương pháp chiết xuấtMagnetic bead method, magnetic rod comb extraction
Số mẫu trên mỗi lần chạy1 để 48
Maximum Reaction System1 mL
Khối lượng xử lý50 µL đến 1000 µL
Khối lượng rửa giải40 µL đến 200 µL
Thời gian khai thác15 to 30 min for 48 samples (reagent dependent)
Sample Pipetting TimeAbout 13 min for 48 samples
Độ đồng đều của chiết xuấtCV ≤ 5%
Single-Sample ConsumablesHỗ trợ
Reagent FormatBottled or prefilled cartridges
Nguyên lý hút pipetDịch chuyển không khí
pipetTwo 1000 µL pipettors and one 200 µL pipettor
Phạm vi hút pipet1 µL đến 1000 µL
Phát hiện chất lỏngPressure detection: liquid level, tip clogging, residual liquid
Lời khuyênDedicated tips and compatible third-party tips
nhiệt độ làm nóngLysis and elution control: room temperature to 120 °C
Khối làm mát0 °C to 105 °C (optional)
PCR Setup Capacity2 × 96-well PCR plates or 24 × 8-strip PCR tubes
Giá đỡ mẫu48 positions for sample tubes, test tubes, and centrifuge tubes, with barcode scanner
Quản lý thông tinAutomatic barcode scanning; external USB scanner support
Tiệt trùng bằng tia cực tímSeparate UV lamps for extraction area, pipetting area, and waste container area
Bộ lọc quạtThree HEPA filters, negative pressure system
Instrument PortUSB
điện áp220 V, 8 A, 50/60 Hz (110V is also available)
Công suất tiêu thụ30 W standby, 550 W maximum
Tiếng ồn: ≤ 60 dB
Kích thước (W × D × H)1090 × 695 × 724 mm
Khối lượng tịnh130 kg

Pipetting Performance

Tham số200 µL Pipettor1000 µL Pipettor
Phạm vi hút pipet1 µL đến 200 µL5 µL đến 1000 µL
tính chính xác1 µL: ±12%; 20 µL: ±3%; 100 µL: ±1%; 200 µL: ±1%5 µL: ±5%; 100 µL: ±2%; 500 µL: ±1%; 1000 µL: ±1%
Độ chính xác (CV)1 µL: ≤ 8%; 20 µL: ≤ 1.5%; 100 µL: ≤ 0.8%; 200 µL: ≤ 0.5%5 µL: ≤ 2.5%; 100 µL: ≤ 1.25%; 500 µL: ≤ 0.4%; 1000 µL: ≤ 0.4%

Bố cục boong

Area Mô tả Chi tiết
Pipetting Arm3-channel arm with two 1000 µL pipettors and one 200 µL pipettor
Khu vực mẫu48-position rack for sample tubes, test tubes, and centrifuge tubes, with barcode scanner
Reagent AreaSeparate zones for extraction reagents and PCR reagents
Khu vực đầuTwo 1000 µL and one 200 µL tip racks
Khu rác thảiTip waste container with fill-level calculation
Magnetic Bead VortexBead resuspension module
Khu vực chiết tách axit nucleic48-sample extraction with heating incubator up to 120 °C
PCR Setup Area2 × 96-well PCR plates or 24 × 8-strip PCR tubes

Hàng tiêu dùng

Mục Mô tả Chi tiết
96 Deep-Well Plate2.2 mL plate for extraction reagents
Magnetic Rod CombDisposable comb that covers the magnetic rods
5 mL Sample TubeTube for sample loading
1.5 mL and 7 mL Centrifuge TubesTubes for samples and reagents
15 mL Wide-Mouth BottleBottle for bulk reagents
100 mL ReservoirReservoir for high-volume reagents
200 µL and 1000 µL TipsBoxed tips for the pipetting arm

NHẬN BÁO GIÁ